Medwarm Neonatal Warming Mattress Cleaning Instructions

This contains all of the cleaning information for the Medwarm Neonatal Warming Mattress.

Medwarm Neo Feature Image 1

Cleaning and Disinfection

  • Control units, mattresses, and blankets must be disinfected before and after use.
  • Disconnect all power and cables before cleaning.
  • Use colourless, residue-free disinfectants suitable for medical surfaces.
  • Compatible disinfectants include 70% ethanol, Fugaten spray derivatives, and chlorine-based solutions with concentration of <10,000ppm.
  • Do not immerse, autoclave, or machine-wash the units or pads.

Control Unit:

The entire surface, cable and control unit should be wiped with a cloth, moistened with suitable disinfectant - do not spray directly onto the control unit. Then wipe with a cloth dampened with water before drying, to remove any residue.

Mattress/Blanket:

Moisten a cloth with suitable disinfectant and clean all surfaces. After disinfection, wipe with a soft cloth dampened with water to remove residues and then pat dry. Failure to rinse and dry may affect the longevity of the mattress/blanket cover and invalidate the warranty.

For more detail, please refer to BHTA guidance on the care, cleaning and inspection of healthcare mattresses: BHTA-3.pdf.

Cross-Contamination Prevention

  • Disinfect all components between patients. Keep pads dry throughout use.
  • Do not allow accumulation of antiseptic or iodine-based fluids beneath patients.
  • Inspect for fluid ingress; do not use if contamination is suspected.

Handling and Storage

  • Store pads flat, dry, and away from sunlight and humidity. Do not fold or compress excessively.
  • Avoid placing sharp or heavy objects on stored pads.
  • Mattress should be inspected before each use to ensure integrity.