Hot Cot Cleaning Instructions
This contains all of the cleaning information for the Croyde Hot Cot.
Warming Mattresses
- Control units and mattresses must be disinfected before and after use.
- Disconnect all power and cables before cleaning.
- Use colourless, residue-free disinfectants suitable for medical surfaces.
- Compatible disinfectants include 70% ethanol, Fugaten spray derivatives, and chlorine-based solutions with concentration of <10,000ppm.
- Clinell wipes are also compatible and can be used to clean the mattress.
- Do not immerse, autoclave, or machine-wash the units or pads.
Cots
-
Wipe clean after each patient use.
-
Check for visible dirt / contamination and surface damage every day.
-
The unit is designed for repeated use in hospitals.
-
The cot frames can be cleaned with Clinell wipes or disinfectant chlorine solutions from 1,000 to 10,000 ppm. We do not recommend using Clinell wipes or chlorine solutions on the tub as this can affect the material and longevity.